Hantek 2C42

190,00 

In stock

Compare

Hantek 2C42

An all-new handheld oscilloscope-Hantek2000 series, it continues the compact and lightweight concept but with a new generation of 3 in 1 multifunction tester: oscilloscope, signal source and multimeter; A small volume of the non-profit grip by a single hand; The suspendingstand was designed to support multi-angle adjustment; Ultra-low-power, Type C charging interface, each 2 hours battery charge allow working one full day or standbytwo months; High-definition color LCD screen display clear and delicate; Minimalist keyboard design simplifyoperation, easy to use.

 

Model Hantek2D72 Hantek2D42 Hantek2C72 Hantek2C42
Oscilloscope Mode
Bandwidth 70MHz 40MHz 70MHz 40MHz
Channel 2CH+DMM+AWG 2CH+DMM+AWG 2CH+DMM 2CH+DMM
Horizontal
Sample Rate Range 250MSa/s(Single-channel), 125MSa/s(Dual-channel)
Waveform Interpolation (sin x)/x
Record Length Max. 6K for single-channel; 3K samples per dual-channel
SEC/DIV Range 5ns/div~500s/div  1, 2, 5 sequence
Vertical
A/D Converter 8-bit resolution,each channel sampled simultaneously
VOLTS/DIV Range 10mV/div~10V/divat input BNC
Bandwidth Limit, typical 20MHz
Low Frequency Response (-3db) ≤10Hz at BNC
Rise Time at BNC, typical ≤5ns
DC Gain Accuracy ±3% for Normal or Average acquisition mode, 10V/div to 10mV/div
Note: Bandwidth reduced to 6MHz when using a 1X probe.
Acquisition
Acquisition Modes Normal
Trigger
Type Edge
Mode Auto, Normal, single
Level ±4 divisions from center of screen
Trigger Level Accuracy 0.2div × volts/div within ±4 divisions from center of screen
Slope Rising, Falling, Rising & Falling
Source CH1/CH2
Input
Coupling DC, AC or GND
Input Impedance, 25pF±3 pF, 1MΩ±2%
DC coupled
Probe Attenuation 1X, 10X
Supported Probe Attenuation Factors 1X, 10X, 100X, 1000X
Maximum Input Voltage 150VRMS
Measurement
Cursor Voltage difference between cursors: △V
Time difference between cursors: △T
Automatic Measurements Frequency, Amplitude
Arbitrary Waveform Generator Mode
Waveform Frequency Sine: 1Hz~25MHz           –           –
Square: 1Hz~10MHz           –           –
Ramp: 1Hz~1MHz           –           –
EXP: 1Hz~5MHz           –           –
Sampling 250MSa/s           –           –
Amplitude 2.5Vpp(50Ω)           –           –
5Vpp(High impedance)           –           –
Frequency Resolution 0.10%           –           –
Channel 1CH waveform output           –           –
Waveform Depth 512Sa           –           –
Vertical Resolution 12 bit           –           –
Output Impedance 50 Ω           –           –
DMM
Maximum Resolution 4000 Counts
DMM Testing Modes Voltage,Current,Resistance,Capacitance,Diode &On-Off
Maximum Input Voltage AC:600V, DC: 600V
Maximum Input Current AC: 10A, DC:10A
Input Impedance 10MΩ
Measurement Term Range Accuracy Resolution
DC Voltage 400.00mV ± (0.8% + 5) 100uV
4.000V 1mV
40.00V 10mV
400.0V 100mV
600.0V ± (1% + 2) 1V
Overload protection: 400mV: 250V, other: 600Vrms.
AC Voltage 4.000V ± (1.2% + 5) 1mV
40.00V 10mV
400.0V 100mV
600.0V ±(1.5% + 5) 1V
Frequency: 40Hz~400Hz;
Frequency of 400V and 600V: 40Hz~100Hz
DC Current 40.00mA ± (1% + 2) 10uA
200.0mA ± (1.5% + 2) 100uA
4.000A ± (1.8% + 2) 1mA
10.00A ± (3% + 2) 10mA
Overload protection:
self restoring fuse: 200mA/250V, 4A and 10A range no fuse.
AC Current 40.00mA ± (1.3% + 2) 10uA
400.0mA ± (1.8% + 2) 100uA
4.000A ± (2% + 3) 1mA
10.00A ± (3% + 5) 10mA
Frequency: 40Hz~400Hz;
self restoring fuse: 200mA/250V, 4A and 10A range no fuse.
Resistance 400.0Ω ±(1% + 3) 0.1Ω
4.000KΩ ±(1.2% + 5) 1Ω
60.00KΩ 10Ω
400.0KΩ 100Ω
4.000MΩ 1KΩ
40.00MΩ ± (1.5%±3) 10KΩ
Overload protection: 220Vrms
Capacitance 40.00nF ±(3% + 5) 10pF
400.0nF 100pF
4.000uF 1nF
40.00uF 10nF
100.0uF 100nF
Overload protection: 220Vrms
Diode 0V~1.0V
On-Off <50Ω
General Specifications
Display
Display Type 2.8 inch64K color TFT
Display Resolution 320 horizontal by 240 vertical pixels
Display Contrast Adjustable
Power Supply
Supply Voltage 100V-240VAC, 50Hz-60Hz; DC INPUT: 5VDC, 2A
Power Consumption <2.5W
Fuse T, 3A
Battery 2600mA*2
Environmental
Operating Temperature 0~50 °C (32~122 °F)
Storage Temperature -40~+71 °C (-40~159.8 °F)
Humidity ≤+104℉(≤+40°C): ≤90% relative humidity
106℉~122℉ (+41°C ~50°C): ≤60% relative humidity
Cooling Method Convection
Altitude Operating and 3,000m (10,000 feet)
Nonoperating
Mechanical Shock Random Vibration 0.31gRMS from 50Hz to 500Hz, 10 minutes on each axis
Nonoperating 2.46gRMS from 5Hz to 500Hz, 10 minutes on each axis
Operating 50g, 11ms, half sine
Mechanical
Dimension 199 x 98x 40mm (L x W x H)
Weight 624g
Standard Accessories Probe*1,Clip Probe*2, Type Charging Cable*1, Power Adapter*1, Multimeter Probe*1 Probe*1,Clip Probe*2, Type Charging Cable*1, Power Adapter*1, Multimeter Probe*1 Probe*1,Clip Probe*1, Type Charging Cable*1, Power Adapter*1,  Multimeter Probe*1 Probe*1,Clip Probe*1, Type Charging Cable*1, Power Adapter*1, Multimeter Probe*1